Ion Selective Electrode for Silver and Sulfide

Ag+/S2-

Technical Data Sheet

Structure of the electrode

range*:
  • 0.1 to 10,000 mg/L silver
  • 0.03 to 200 mg/L sulfide

The silver- and sulfide - sensitive electrode is a solid-state membrane electrode with a highly compressed Ag2S sensor element embedded in a plastic shaft by means of a special resin.
A silver contact layer and a shielded cable are used for potential discharge.

Ordering Information

PO No. Type Description
116 100 Ag/S selective electrode with 1m sealed cable
116 101 Ag/S selective electrode with 1m sealed cable with BNC connector
116 102 Ag/S selective electrode with 1m sealed cable with DIN connector
116 005 Ag/S selective electrode with indutrial threaded head
117 004 Ag/S selective threaded mini electrode with BNC connector
slope of characteristic:
  • 57 (+-)2 mV/pAg
  • 27 (+-)2 mV/pS
response time in 10-3 M
(107 mg/L silver or 32 mg/L sulfide):
  • < 30 sec
drift within 30 minutes:
  • 0.3 mV
drift within 24 hours:
  • 1 mV
concentration of interfering ion species (relative to silver and sulfide):
  • Hg2+ and OH- affect silver solution (precipitates)
  • H+ affects sulfide solution (formation of H2S)
pH range:
  • 1 - 7 (determinations of silver)
  • 7 - 14 (determinations of sulfide)
temperature range:
  • 0 - 80 °C
resistance:
  • < 0.01 MOhm
sensor material:
  • die-pressed Ag2S, silver contact
shaft material:
  • plastic
dimensions:
  • length 145mm
  • diameter 12 mm
* approximate values, depend upon specific application