Ion Selective Electrode for Calcium

Ca2+

Technical Data Sheet

Structure of the Electrode

range*:
  • 2.5 10-6 to 1 mole calcium
  • 0.1 to 40,000 mg/L calcium

The calcium selective electrode is a matrix membrane electrode with a solid contact contact. The flexible ammonia selective membrane is made from a special organic ion exchanger which is dissolved in an organic solvent and embedded into a PVC matrix.
As an organic phase the resulting "liquid membrane" is fixed by the PVC matrix and renders a stable phase boundary.
An ion selective solid state membrane forms the discharge contact for the equilibrium voltage which develops at the phase boundary.
Both the shaft and the ion selective membrane are made from PVC. This should be taken into account when using the electrode.

Ordering Information

PO No. Type Description
102 100 Ca selective electrode with 1m sealed cable
102 101 Ca selective electrode with 1m sealed cable with BNC connector
102 102 Ca selective electrode with 1m sealed cable with DIN connector
102 005 Ca selective electrode with indutrial threaded head
103 004 Ca selective threaded mini electrode with BNC connector
slope of characteristic:
  • 27 (+-)2 mV/pCa
response time in 10-2mole,
400 mg/L calcium solution:
  • < 30 sec
drift within 30 minutes:
  • 0.3 mV
drift within 24 hours:
  • 1 mV
concentration of interfering ion species:
  • Mg2+: 3 102 times
  • Pb2+: 2 times
  • Zn2+: 1 times
  • Fe2+/ Cu2+: 10 times
  • Sr2+: 0.6 times
  • Ba2+: 0.5 times
pH range:
  • 3 - 9
temperature range:
  • 0 - 40 °C
resistance:
  • < 1 MOhm
sensor material:
  • organic ion exchanger in a special solvent, homogeneously distributed in PVC
shaft material:
  • plastic
dimensions:
  • immersion depth 120mm
  • diameter 12 mm

* approximate values, depend upon specific application