Ion Selective Electrode for Iodide and Cyanide

I-/CN-

Technical Data Sheet

Structure of the Electrode

range*:
  • 0.01 to 5,000 mg/L iodide
  • 0.03 to 260 mg/L cyanide

The iRAS iodide/cyanide selective electrode is a solid state membrane electrode. the ion selective membrane consists of a gighly compressed AgI/Ag2S sensor element embedded in a plastic shaft by means of a special resin.

Ordering Information

PO No. Type Description
106 100 I/CN selective electrode with 1m sealed cable
106 101 I/CN selective electrode with 1m sealed cable with BNC connector
106 102 I/CN selective electrode with 1m sealed cable with DIN connector
106 005 I/CN selective electrode with indutrial threaded head
107 004 I/CN selective threaded mini electrode with BNC connector
slope of characteristic:
  • 57 (+-)2 mV/pI or pCN
response time in 26 mg/L cyanide or 127 mg/L iodide solution:
  • < 30 sec
drift within 30 minutes:
  • 0.3 mV
drift within 24 hours:
  • 1 mV
concentration of interfering ion species:
  • Br-: 3 10-3 times
  • Cl-: 106 times
  • S2-: traces
  • Hg2+, Ag+, Pb2+, Tl+: affect solution to be analysed (precipitation)
pH range:
  • 7 - 14 (cyanide determination)
  • 2 - 14 (iodide determination)
temperature range:
  • 0 - 80 °C
resistance:
  • < 0.1 MOhm
sensor material:
  • AgI/Ag2S die-pressed
shaft material:
  • plastic
dimensions:
  • immersion depth 120mm
  • diameter 12 mm

* approximate values, depend upon specific application